User Guide
- 60
–
20 Troubleshooting
Display
Decimal Point
Status
Alarm
Message
Description
XXXX.
Flashing
Off
L. Hrs Mode
Displays the accumulated hours the lamp has
been on
XXX.X
Solid
Off
Timer Mode
Displays the time in seconds the shutter will
remain open when the start button is pressed
XX.XX
Solid
Off
LEVEL Mode
Displays Irradiance (W/cm2) if the S2000 has
been calibrated. CAL Led is ON.
XXX
None
Off
Iris Opening
Displays the percentage the iris is open while in
Intensity Mode.(S2000 is not calibrated)
Flashing
None/Flash
Off
Flashing
display
The lamp is warming up. Warm up time is
approximately 4 minutes.
bulb
None
On
Bulb Error
Lamp did not strike or extinguished after
striking
cool
None
Off
Cool Warning
Lamp is too hot to strike.
The lamp will automatically re-strike when it has
reached optimum temperature.
old / bulb
None
Off
Alternating Old
and then Bulb
The lamp has accumulated over 2000 hours.
Lamp may be near end of life.
end / bulb
None
On
Alternating
End and then
Bulb
The lamp has accumulated over 4000 hours.
The lamp will not strike.
SF1
None
On
Shutter Failure Shutter has failed. Unit should be serviced
CA1
None
Off
Calibration
S2000 unit is doing calibration
On
Calibration
Failure
The calibration has failed
LOC
None
Off
Adjustment
Locked
The up/down adjustment buttons have been
locked. No changes can be made to the
exposure time or iris setting.
ULOC
None
Off
Adjustment
Unlocked
The up/down adjustment buttons have been
unlocked. Changes can be made to the
exposure time and iris setting.
E1
None
Off
Wrong Type of
Lamp Installed
Check lamp part# to verify that the appropriate
lamp type has been installed.
E3
None
On
Lamp
exceeded
temperature
The lamp temperature monitoring has turned off
the lamp due to excessive heat. Verify the
cooling exhaust vent is not blocked.
E4
None
On
Internal error.
An internal error has occurred. If error re-occurs
the unit should be serviced
E5
None
On
Failed to
communicate
with intellilamp
The communication to the intellilamp has failed,
open the lamp access door and verify all
connections are secure.
EF1
None
On
Exposure
Failure
Selected output intensity was not delivered.
CLF
None
Off
Intensity Mode
Relative
S2000 is set to relative Intensity Mode. Shutter
is closed. (S2000 is not calibrated). Closed-
Loop Feedback is available.
Table 10 Message References
Note:
The Alarm will be “On” if the light guide is not fully inserted and the shutter is supposed to be open