User Guide
- 57
–
19.1 Replacing the Lamp Module
19.1.1
The S2000 comes configured for either standard or surface curing applications.
Please ensure that the appropriate type lamp has been selected for your curing
system/ application. The part number of the lamp module is available on the
lamp box.
Lumen Dynamics Part #
Description
012-64000R
Standard Curing Lamp
012-65000R
Surface Curing Lamp
Table 8 Lamp Part Number
Refer to Section
4
for lamp module replacement instructions
Refer to Section
20
for messages regarding troubleshooting and lamp module
replacement.
19.2 Replacing the External Fuses
19.2.1
The external (mains) fuses are located in the fuse drawer which is located in
the AC inlet module on the rear panel.
19.2.2
Turn off the main POWER switch and remove the AC POWER cord from the
unit.
19.2.3
Gently pull out the drawer with the aid of a flat-head screwdriver.
19.2.4
Carefully lever one end of the blown fuse up from its retaining clip with a small
flat-head screwdriver and lift it out.
19.2.5
Replace the damaged fuse(s) only with the same type and rating (F5A, 250V).
The rear compartment must contain two active fuses.
19.2.6
Close the fuse drawer.
19.2.7
Reconnect the AC POWER cord.
19.3 Replacing the Air Filter
19.3.1
The external air filter is located under the front face plate of the S2000.
19.3.2
Turn off the main POWER switch and remove the AC POWER cord from the
unit.
19.3.3
Gently slide and pull out the filter.
19.3.4
Push in the replacement filter so that it sits flat in place