6 Powering Up and Powering Down
6.1.1
Ensure that the lamp and light guide have been properly installed and that the
lamp housing panel is securely fastened.
6.1.2
Plug the S2000 unit into a properly grounded AC outlet.
6.1.3
Turn on the mains POWER switch “I “, located on the front panel and check the
fan for airflow.
6.1.4
As soon as the LED display turns on it will display the version of software
currently programmed in the S2000 (i.e. R
– X). Once the software version
level has been indicated, the display will reset after several seconds to the next
display mode.
6.1.5
The lamp will automatically turn on within 45 seconds and the lamp indicator
will illuminate and flash. The front panel display will flash during the warm up
period for approximately 4 minutes. The display will stop flashing when the
warm up period has completed.
,
Lumen Dynamics
recommends 20 minutes to
ensure a stable optical output
.
Warning
: If the lamp is turned off, and an attempt is made to turn it back on
before it has fully cooled, the “cool” message will appear on the
display. The lamp will automatically re-strike when the lamp has
cooled.
6.1.6
To power down the unit, set the mains power located on the front panel to “0”.
6.1.7
10 exposures must be run to save any new exposure settings (intensity &
timer) into memory for it to remember next time the system is powered-up.
Lamp Warm-UP:
The ARC lamp has 3 distinct phases of operation;
1.Ignition.
2
.
Warm-up.
Lumen Dynamics
recommends 20 minutes of proper warm-up
and to ensure a stable optical output.
3. Stable Operation.
It is recommended that phase 1 and 2 are not interrupted. This can result in shortened
lamp life.
The lamp must be allowed to warm-up uninterrupted.