![Lumen Dynamics Group OmniCure S2000 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/lumen-dynamics-group/omnicure-s2000/omnicure-s2000_user-manual_1947998016.webp)
User Guide
- 11
–
5 Inserting and Removing the Light Guide
5.1.1
Ensure that the protective end caps are removed from both the input and
output ends of the light guide prior to installation.
5.1.2
Insert the light guide into the light guide port located on the front panel of the
unit. Push the light guide in until it seats with a second positive "click".
5.1.3
During installation or removal, grasp the light guide on the strain-relief nearest
the input end of the light guide.
5.1.4
To remove the light guide, firmly grip the strain relief near the light guide
retainer and pull out firmly.
5.1.5
Refer to section 19.5 for Light Guide Cleaning Instructions.
Note:
3mm single leg liquid light guides are not compatible with the
S2000!
Tip:
When the light guide is fully inserted, the LED above the light
guide port will illuminate green when POWER is on to the unit. If
the light guide is not fully inserted the LED will illuminate red.
Note:
the shutter will not open if the light guide is not fully inserted.
Note:
Never grip the light guide during installation or removal in a place
other than the strain relief portion of the light guide.
Note:
The S2000 is designed for use with
Lumen Dynamics
Light Guides.
Lumen Dynamics
can not guarantee the performance of the S2000 if
using light guides other than those supplied by
Lumen Dynamics
.