18.3 PLC Shutter Mode
18.3.1
To change PLC Shutter Signal mode, you need to open the administrator
screen, go to the Administrator menu and click on Setup System.
The following is the administrator screen:
The PLC Shutter Signal Mode can be switched between the pulsed
start/stop and the level start/stop without password.
18.3.2
The pulsed start/stop opens and closes the shutter when the S2000 shutter
contacts are closed. The shutter will automatically close on the next closure of
the contacts if it is open, if it is closed it will start the next timed exposure.
18.3.3
The level start/stop mode opens the shutter when the shutter inputs are closed
and closes the shutter when the shutter inputs are opened. When PLC Shutter
signal mode is in Level Start/Stop the S2000 must be also in level mode for the
shutter to open/close with the trigger. The minimum exposure time in Level
Mode is 0.5s.
Note:
The shutter will not respond to the PLC inputs if it has been opened by using
the front panel start/stop button or if the shutter open command has been
issued from the Control Panel.