28. Set irradiance set point
Command to S2000;
SIRn
Note:
n > 0, and n is a float number
Example:
If
“SIR15.32XX\r” is set, the S2000 irradiance set point
will be 15.32 W/cm2.
Response from S2000;
Received
– if the value n is valid
Invalid
– if the value n is out of range
Need CAL
– the S2000 need to be calibrated before the
value can be set.
EXP
– the S2000 is running a timed exposure, so the
value cannot be set.
29. Get PLC Mode
Command to S2000;
GPM
Response from S2000;
n
Note
: If n is „1‟ then trigger level mode is enabled, if it is „0‟
then trigger level mode is disabled.
30. Set PLC Mode
Command to S2000;
SPMn
Note
: n, if n is 0 then trigger level mode is disabled. If n is
1 then trigger level mode is enabled.
Response from S2000;
Received
– if the command has been accepted.
Invalid
– if the value is outside the specified range.
16.5 Command Timing Specification:
16.5.1
The minimum amount of time required for data to be received by the PC is over
20mS except for any commands related to iris movement which is over
200msec. The next command must be sent after the previous command is
responded by the unit. Failure to adhere to this requirement will result in
dropped responses as well as other undesirable effects.
16.6 Graphic User Interface Control:
16.6.1
When the unit is connected to PC, the unit front panel will be automatically
locked out. The unit status will be sampled at twice per second. Refer to
section 20 for detailed operating description of the GUI.