User Guide
- 36
–
16.4 Command Descriptions:
1. Connect S2000
Command to S2000;
CONN
Response from S2000;
READY
2. Disconnect S2000
Command to S2000;
DCON
Response from S2000;
CLOSE
3. Lock the front panel
Command to S2000;
LOC
Response from S2000;
Received
4. Unlock the front panel
Command to S2000;
ULOC
Response from S2000;
Received
5. Clear Alarm
Command to S2000;
CLR
Response from S2000;
Received
6. Get unit status
Command to S2000;
GUS
Response from S2000;
n
(n represents a integer number)
Example:
If “50XX\r” is received, the unit status is as follows:
Alarm - Off; Shutter Home
– Pass; Lamp – On; Lamp
Ready
– Yes; Shutter – Close; Front Panel – locked;
Calibration
– No; Exposure Fault – No.
7. Run a timed exposure
Command to S2000;
RUN
Response from S2000;
Received
8. Open Shutter
Command to S2000;
OPN
Response from S2000;
Received
9. Close Shutter
Command to S2000;
CLS
Response from S2000;
Received
Note:
Please refer to Table 7 Unit Status for details.