13 Messages / Indicators
The S2000 display and LED indicators provide information to the user to aid in unit
operation and to advise when certain conditions exist. The following is a collective
listing of messages and their meanings.
1. "XXXX."
A flashing decimal point to the right of the four numbers on the display advises the
user that the unit is in Lamp Hour Mode. See Section
12,
Viewing the Accumulated
Lamp Hours.
2. "XXX.X"
A decimal point to the left of the furthest right digit of the four digits on the display
indicates the unit is in Timer Mode. The number shown will indicate the exposure
time in seconds and tenth of seconds. See Section
10
, Timed Exposures.
3.
“XX.XX”
The unit displays Irradiance in W/cm2 if it has been calibrated and set to Absolute
Intensity Mode. Decimal point is solid.
4.
“CLF”
The unit has not been calibrated and has been set to Relative Intensity Mode.
Shutter is closed. Closed-loop feedback is available.
5. "XXX"
Three digits on the display indicate that unit is in Intensity Mode and shutter is open.
The three digit number shown will indicate the percent iris opening. See Section
8
,
Adjusting the Light Output.
6. "bulb"
The "bulb" message advises the user that the bulb has failed to ignite or
extinguished after striking.
If the bulb message appears immediately after power up and is accompanied by an
audible beeping, this indicates that the lamp has not been installed correctly. See
Section
4
, Installing the Lamp Module.
If the bulb message appears after about 45 seconds and is accompanied by an
audible beeping, this indicates that the lamp has failed to strike. It may be a result of
the lamp reaching end of life, that the lamp housing panel is not secured properly in
place or that the lamp power connector is not plugged in. Press the start/stop button
to clear the audible alarm. Turn power off to the unit and disconnect power cord.
Check that the lamp housing panel is secured properly in place and verify that all
lamp connections as correct as per section
4
. Wait a few minutes and turn power on
to the unit. If it still does not strike, re-check all electrical connections to the lamp.