61
ULM.0.01.000RE_10_08_20
Failure
Cause
Remedy
Signal
4-20mA corre-
sponds to one
of alarm val-
ues
One or more controlled pa-
rameters of transmitter are
beyond the permissible val-
ues:
Connect to transmitter through PC
or Bluetooth. Check whether alarm
signal value matches its mask (para.
6.4).
- transmitter temperature is
beyond the range
Remove overheat cause, insulate
level transmitter from increased
temperature process by means of ra-
dio-transparent shield (see 12.5), use
transmitter forced-air cooling, install
sun shield.
- lack of transmitter perpen-
dicularity (for application
with liquid products)
Correct by device leveling.
- unreliable temperature or
tilt angle measurement
Related channels in the device elec-
tronic module are faulty. Further op-
eration is allowed subject to over-
heat prevention. Otherwise, return
device for repair.
- unstable readings due to
product surface disturbance,
unstable surface.
Indicative parameter, further opera-
tion is allowed.
- transient conditions – there
is no stable reflecting prod-
uct surface, or transmitter is
at the initialization stage
Transmitter in initial switching-on
mode; device will exit this mode in
1-2 minutes after the switching on.
Otherwise, record spectrum (see pa-
ra. 7.3.), send it to the technical sup-
port address.
- measuring device mal-
function – level transmitter
electronic unit failure
Return device for repair.
Содержание ULM-31-HF
Страница 28: ...28 ULM 0 01 000RE_10_08_20 Fig 4 12 Examples of Installation ...
Страница 74: ...74 ULM 0 01 000RE_10_08_20 12 5 Mounting of ULM 31А1 level transmitter at the stub tube ...
Страница 76: ...76 ULM 0 01 000RE_10_08_20 12 7 Mounting of ULM 31 HF ULM 31А1 HF level transmitters at the stub tube ...
Страница 77: ...77 ULM 0 01 000RE_10_08_20 12 8 Mounting of ULM 31А1 level transmitter at the stub tube with insulating spacer ...
Страница 83: ...83 ULM 0 01 000RE_10_08_20 12 14 Mounting flange ...