ULM.0.01.000RE_10_08_20
4.6 Examples for Installation
The examples for the level transmitter installation designed in accordance with this
instruction requirements are shown in Fig. 4.12 in positions А and B. Installation of
the level transmitter on the tank flange without using the pipe branch is shown in
Fig. 4.12 Variant А and installation using the pipe branch - on Variant В. Variants
C-Е show the installation errors, which should be paid attention to.
Variant C – the roof edge under the pipe branch extends inside the pipe branch. Be-
sides, the opening in the roof is smaller than the minimum permissible one.
Variant D - the pipe branch lower edge is embedded in relation to the roof, due to
which the pipe branch length is longer the permissible one.
Variant Е - the level transmitter is installed with high displacement in relation to the
pipe branch vertical axis. The device should be installed according to the pipe
branch central axis.