19
ULM.0.01.000RE_10_08_20
- to provide the level measurement over the total depth of the tank, the level trans-
mitter shall be oriented in direction of the tank lowest point. In the vertical cylindri-
cal hoppers with the cone-shaped outlet, this is achieved by its installation in the
center of the hopper roof.
Fig. 4.4. Device Installation on Hopper with Cone-Shaped Bottom.
In case such way of installation is impossible, it is necessary to direct the device
measuring beam to the hopper center, which is achieved by inclining the mounting
pipe branch flange plane.
The required inclination angle
depends on installation position and the hopper
dimensions. The inclination angle can be checked using the inclination angle sensor
built in the level transmitter or using the level (For details, see para. 7.1 “Device
Diagnostics”).