Missing or damaged sealing ring impairs the device ingress protection and can
result in its failure.
4. Strip the cable of external insulation. When using shielded cable, remove
shield.
5. Strip wires of insulation 4-6mm from end. When using multiconductor wires,
it is recommended to crimp bare ends with lugs.
6. Connect the cable in accordance with the pin assignment diagram. According
to para. 5.4 – 5.5 and following instructions on device nameplate. Tighten screw
terminals with 3mm screwdriver.
Incorrect connection can result in device failure.
Fig. 5.12 Level transmitter connection
7. Slightly pull wires in order to check whether these are securely fixed in the
terminal block contacts.
8. Adjust cable length as necessary for connecting to terminals and tighten cable
gland nut securely. Sealing ring must fit tight to cable circumference.
9. Check cover sealing ring (para. 12.15, pos. 3) availability and integrity and
screw device cover on.
10. Connect external grounding terminal to vessel grounding using 6mm screw-
driver.
!
CAUTION
!
CAUTION