Display and control panel AMS 348i SSI
Leuze electronic GmbH + Co. KG
AMS 348
i
SSI
49
TNT
35/7-2
4V
SSI submenu
Position value submenu
Note!
The parameters mentioned in the "Position value", "I/O" and "Other" submenus must be set
via the GSDML file of the AMS 348
i
. For test purposes, the parameters can be changed
directly on the AMS 348
i
via the display structure described below. After deactivation of the
parameter enable, however, the parameters stored in the GSDML file continue to be valid.
The parameters changed directly on the AMS 348
i
for test purposes are overwritten again.
q~ÄÉääÉ=UKPW
ppf=ëìÄãÉåì
Level 3
Level 4
Level 5
Selection/configuration option
Description
Standard
Activation
ON/OFF
Activates or deactivates the AMS 348
i
SSI as an SSI partic
-
ipant.
ON
Encoding
Binary/gray
Specifies the output format of the measurement value.
Gray
Number of
data bits
24-bit/25-bit/26-bit
The measurement value can be represented on the SSI inter
-
face in this data width.
24 bit
SSI resolution
0.001mm / 0.01mm / 0.1mm / 1mm / 10mm / free resolu
-
tion
The measurement value can be displayed in these resolu
-
tions.
The value of the free resolution is determined in the "Position
value" submenu in the "Free resolution value" parameter.
0.1mm
Error bit
ON/OFF
This parameter determines whether an error bit is also
attached to the "number of data bits". The error bit is the LSB
and is not converted in the case of gray representation of the
measurement value.
ON
Error bit func
-
tion
The error bit can be assigned the following status messages:
Overflow / Intensity (ATT) / Temperature (TMP) /
Laser (LSR) / Plausibility (PLB) / Hardware (ERR).
In the case of multiple entries, the individual states in the error
bit are processed in an OR function.
Plausibil
-
ity (PLB)
Hardware
(ERR)
Clock fre
-
quency
50 - 79kHz / 80 - 800kHz
Selection of the clock frequency.
80 -
800kHz
Table 8.4:
Position value submenu
Level 3
Level 4
Level 5
Selection/configuration option
Description
Standard
Unit
Metric/Inch
Specifies the units of the measured distances
Metric
Counting
direction
Positive/Negative
Positive: The measurement value begins at 0 and increases
with increasing distance.
Negative: The measurement value begins at 0 and decreases
with increasing distance. Negative distance values may need
to be compensated with an offset or preset.
Positive