25
PROCEDURE Setting Tilt Correction
Operating Procedure
Operation
Display
⑴
Turn on the machine, enter MEAS screen.
POWE
R
MEAS
.
PC -30
PPM
0
3
ZA 92°36′25″
HAR 120°30′10″
P1
SD SHV HSET EDM
⑵
Press ESC to enter status screen.
ESC
2021-01-01 10
:
00
:
48
KTS-442UT
No.S109996
Ver.2020-1.02
File:JOB01
MEAS MEM CNFG
⑶
From the status screen press CNFG to enter
the configuration setting screen.
CNFG
Config
1. Obs. Condition
2. Instr. Const
3. Date& time
4. Comms setup
5. Unit
⑷
Select “1. Obs. Condition” and press ENT
(or press numeric key 1). Use ▲or▼ key to align
the cursor to the fourth line “Tilt crn”, use
or
to set the tilt correction method. Then
press ENT to finish set. There are three options
about tilt correction: There are three options:
NO, 1-axis and 2-axis.
“1. Obs.
Condition”
+
ENT
+
▲or▼
+
or
Condition
C&R crn.: No
V. obs : Zenith
Tilt crn. : Yes (H& V)
Dist mode: HD
Power off : Off
↓
⑸
Press ESC to return to setting screen.
ESC
Config
1. Obs. Condition
2. Instr. Const
3. Date& time
4. Comms setup
5. Unit
6. Key function
If the instrument is not leveled, it is possible to see different results in different tilt correction settings.
Steps Leveling instrument