172
4.
2C
=L-R±180°=-30″≥±
20
″, adjustment is necessary.
·
Adjustment
A
:
Adjustment by on-board program
:
Operation procedure
KEY
Display
⑴
After levelling the machine, power on, press
CNFG under date/ time screen, as right diagram
shows.
CNFG
2020-09-10 10
:
00
:
48
Type
:
KTS
-
442UT
No.
:
S112926
Ver.
:
20.09.10
Job
:
JOB01
MEAS MEM CNFG
⑵
Press ▼ key to choose“2. Instr. Const.”and press
ENT key(or press numeric key 2),enter the
instrument constant setting screen.
“
2
. Instr.
Const.”
+
ENT
1. Obs. condition
2. Instr. Const.
3. Date& time
4. Comms setup
5. Unit
6. Key function
⑶
press ▼ key to choose “3. collimation”, then
press ENT key(or press numeric key3), enter
collimation error adjustment function.
“3.
collimation”
+
ENT
1. V0/Axis const.
2. V0/ Adjustment
3. Collimation
4. horizontal axis
5. Instr. Const.
6. Contrast ADJ.
⑷
At positive position (Face left) sight at target,
press OK.
Face left,
sight at target
+
OK
Collimation
<Step-1> Front
ZA 0°21′39″
HAR 185°47′57″
OK
⑸
Rotate telescope. At the reverse position (face
right) sight at the same target precisely, press OK.
Face right
and sight at
target
+
OK
Collimation
<Step-2> Reverse
ZA 179°38′17″
HAR 5°50′57″
OK
⑹
After adjustment, it shows “set” on the screen,
the display returns to Instrument constant menu
screen.
1. V0/Axis const.
2. V0/ Adjustment
3. Collimation
4. horizontal axis
5. Instr. Const.
6. Contrast ADJ.