39
repeat step 4 and step 5.
·
When the repetition measurement is completed,
press ESC.
·
In repetition measurement mode, even if “Automatic Tilt Compensation On” is selected, horizontal angle
compensation will not occur.
·
Maximum measurement frequency
:
10 times
·
Repetition display range
:
3599°59′59.5″
8.5 Slope in
%
·
KTS-442UT allows to displays the gradient as a %.
·
In advance allocate the function keys to display [ZA/%]. Refer to “20. Key Function Allocation”.
PROCEDURE
Operating Procedure
Operation
Display
⑴
In MEAS mode, display the screen in which
ZA/℅ is registered.
Allocate
the
function
key to
display
ZA/℅
Meas
.
PC -30
PPM 0
3
VA 89°59′54″
HAR 17°31′50″
P1
SD SHV HSET ZA/%
⑵
Press ZA/℅, the vertical angle (ZA) is
displayed as a gradient (V%)
Press ZA/℅ again to return to the original
vertical angle display.
ZA/℅
Meas
.
PC -30
PPM 0
3
V℅ 0.00 ℅
HAR 17°31′50″
P1
SD SHV HSET ZA/%
☆
Display range : within ±100
%
☆
When parameter “Vertical angle format” is set to “Horizontal 0°” or “Horizontal
0°±90°”, “ZA” is displayed instead of “VA”.
9. DISTANCE MEASUREMENT
·
This section explains the following information about distance measurement. First complete the preparations in
9.1 to 9.2 in order to perform distance measurements.
9.1 Settings for Distance Measurements