37
Lubricants, O-Rings
For retractable fittings used in the chemical industry, the lubricant Syntheso Glep1 (silicone-free) is
applied. For retractable fittings used in the pharmaceutical / food industry (when FDA conformity
is required), the lubricant Beruglide L (silicone-free) is applied (registered according to NSF-H1).
On request, the lubricant Paraliq GTE 703 can be applied (good lubricating properties also at
increased temperatures and for a large number of stroke movements). This lubricant contains
silicone and is only used as special application on specific request.
Application
Pharma / Food
Chemistry / Wastewater
Lubricant
Beruglide L
(silicone-free)
FDA compliant
NSF-H1 registered
Paraliq GTE 703
(containing silicone)
FDA compliant
(USDA H1)
Syntheso Glep 1
(silicone-free)
Materials of
elastomeric gaskets
FKM
–
–
+
FFKM
–
–
+
EPDM
–
–
+
FKM - FDA
+
+
–
FFKM - FDA
+
+
–
EPDM - FDA
+
+
–
+ means: suitable; – means: not suitable
Maintenance
SensoGate WA 133
The drive unit must be removed, for example:
• for general maintenance or inspection
• to clean the calibration chamber, e.g., after a sensor has broken
• to change the sensor / calibration-chamber gaskets
• in the event of a technical fault of the drive unit
Servicing the Drive Unit
Selected Cleaning Agents for Specific Applications
Application
Cleaning agent
Specification
1)
Deposits and dirt
Water + brush
cold or hot
Limescale
Dilute acids
e.g., hydrochloric acid or sulfamic acid
Fat
Surfactant
-
Dilute alkali
e.g., sodium hydroxide
Alcohol solubles
Solvant
e.g., ethanol or isopropyl alcohol
Proteins
Pepsin/hydrochloric acid solution -
1)
Observe the chemical resistance of the retractable fitting!
Содержание SensoGate WA 133 Series
Страница 43: ...43 Annotations SensoGate WA 133 ...