28
Immersion Tube
SensoGate WA 133
1. Push the O-ring
(D)
(sensor gasket) onto
the sensor
(A)
as shown.
2.
Note:
Make sure that there is no further
O-ring in the immersion tube (T)
(installed by mistake).
3. Loosen the two screws
(Z)
by approx.
4 turns (do not detach them) if you
have not done that when removing the
immersion tube.
4.
Note:
If the screws have been screwed
out too far, the immersion tube cannot
be installed (correct if required).
5. Push the immersion tube
(T)
in direction
of the arrow and insert it in the bayonet
coupling
(Y)
.
Note:
The bayonet coupling is locked by the
form-fit screw heads. The immersion tube,
however, remains movable to compensate for
tolerances.
Z
D
A
T
Y
T
Z
6. Press the tube in place and turn it clock-
wise until the stop (approx. 60°).
7. Fasten the two screws
(Z)
using a screw-
driver (TX 25).
Conditions:
The retractable fitting must be in PROCESS position (see “PROCESS Position” chapter).
Installing the Immersion Tube
Содержание SensoGate WA 133 Series
Страница 43: ...43 Annotations SensoGate WA 133 ...