NOTICE: Always install both hoses to ensure safe operation!
To ensure safe operation of the retractable fitting, you must connect the inlet and outlet hoses
and make sure that the cleaning and calibration solutions are collected!
The limit switch converts the pneumatic limit signals into electric output signals (PE converter).
Through a piston, the pressure acting on the push-in fitting actuates the electrical micro push
button switch (normally-open contact) located in the limit switch.
14
Assembly
SensoGate WA 133
1. Connect the outlet hose
(AF)
using the hose cou-
pling
(AE)
. The outlet is
marked with the following
symbol
(I)
.
2. Connect the inlet hose
(AZ)
using the hose coupling
(AE)
. The inlet is marked
with the following symbol .
AE
AF
Mounting the Retractable Fitting
• Possible mounting angle 15° above
horizontal
• Mounting angle 360° (i.e. even upside down)
for special sensors only containing thickened
electrolytes which thus cannot flow.
15°
15°
Installing the Outlet and Inlet Hoses
AZ
AE
I
Installing the ZU 0859 Electrical Limit Switch with Plug-In Adapter (Optional)
1. Screw the plug-in adapters onto
the limit switches.
2. Insert the assembled limit switches
into the
“SERVICE feedback”
and
“PROCESS feedback”
push-in fittings.
“SERVICE feedback”
push-in fitting
“PROCESS feedback”
push-in fitting
Plug-in adapter
Limit switches