Calibration Chamber
SensoGate WA 133
1. Screw off the outlet
(AM)
and
inlet connectors
(AV)
. Take off
the loose flange if required.
2. Loosen and remove the two
screws
(AK)
from the calibration
chamber (using screwdriver
TX25).
K
P
N
3. Pull the calibration chamber
(K)
vertically out of the process
adaptation
(P)
.
4. Push the flange bushing
(N)
downwards out of the process
adaptation.
Now, the gaskets are accessi-
ble and can be checked and
replaced if required.
AK
AK
Removing the Calibration Chamber
AM
AV