NOTICE: No Process Pressure!
Make sure that the fitting is disconnected from process pressure!
Take appropriate safety precautions against escaping process fluids.
24
Drive Unit
SensoGate WA 133
Note:
Follow the steps below in the correct order.
1. Move the retractable fitting to the SERVICE position.
2. Make sure that no process fluid is leaking from the outlet
(I)
.
3. If required, remove the sensor as described (see “Sensors” chapter).
4. Separate the outlet and rinse connection if required.
5. Carefully turn the coupling nut
(AH)
counterclockwise
(using the ZU 0680 accessory wrench
(MS)
if required – see figure).
Take care to not cant the unit!
6. Pull off the drive unit
(AD)
upwards to separate it from the process adaptation
(P)
.
I
AH
AD
MS
Removing the Drive Unit
Mounting wrench,
ZU 0680
P