19
Sensors with Solid Electrolyte
SensoGate WA 133
Short Immersion Depth
A
J
C
D
E
F
G
Removing the Sensor
Only remove the sensor when the retractable fit-
ting is in
SERVICE position
(see “SERVICE Postition”
chapter).
1. Remove the protective cap (ZU 0759) if
required (see user manual of protective cap).
2. Remove the cable jack
(G)
.
3. Before removing the sensor, check that there is
no liquid leaking from the outlet (process seal-
ing might be defective).
4. Remove the sensor (recommended tool:
19 mm mounting wrench, e.g., Knick ZU 0647).
5. Check whether the slide washer
(C)
or the
O-ring
(D)
are damaged.
!
NOTICE: Glass Breakage!
When replacing damaged sensors (glass bro-
ken), be sure to check the sensor gasket in the
immersion tube and replace it if required (see
“Immersion Tube” chapter).