23
Sensors with Liquid Electrolyte
SensoGate WA 133
Removing the Sensor
Before removing the sensor, make sure that the
retractable fitting is in
SERVICE position
(see
page 17 for information on installing or removing
a sensor).
1. Remove the cable jack.
2. Before removing the sensor, check that there
is no liquid leaking from the outlet (process
sealing might be defective).
3. Loosen the small coupling nut
(R)
– do not
remove it.
4. Unscrew and remove the large coupling nut
(S)
and pull the detached unit upwards.
5. Carefully pull out the sensor
(A)
.
R
S
A