Drive Unit
SensoGate WA 133
Note:
Follow the steps below in the correct order.
1. Insert the drive unit into the process adaptation
(P)
(in SERVICE position).
The radial position of the drive unit is determined by a coding pin
(O)
in the calibration
chamber and an opening
(AJ)
in the drive unit.
The coupling nut can only be tightened when the drive unit is in the correct position.
2. Now tighten the coupling nut
(AH)
(turn clockwise – hand-tight or 10 Nm –
using the ZU 0680 accessory wrench
(MS)
if required).
3. If required, remove the sensor as described (see “Sensors” chapter).
P
O
AH
AJ
MS
Installing the Drive Unit