Installing and Removing a Sensor
SensoGate WA 133
Preparations:
• Move the retractable fitting to the SERVICE position.
• If provided, twist the
SensoLock
ring to '
lock
' position.
• Limit the pressure on the fitting to a maximum of 8 bar.
• Make sure that there is no liquid leaking from the outlet.
• Check whether the sensor is damaged (e.g., glass broken).
Never install a damaged sensor!
• Check whether the slide washer or O-ring on the sensor are damaged and replace if required.
• Remove the watering cap from the sensor tip and rinse the sensor with water.
• When the sensor has a silicone seal on the junction (as transport protection),
remove this seal using a knife.
• Follow the assembly instructions given in this manual step by step.
!
NOTICE: Sensor Installation or Removal by Qualified Personnel Only!
Sensors shall only be installed or removed by personnel authorized by the operating company
and trained in the handling of the retractable fitting.