E
F
22
Sensors with Liquid Electrolyte
SensoGate WA 133
Note:
Observe the user manual of the sensor.
In the case of inclined installation, turn the elec-
trolyte filling hole
(W)
of the sensor towards
the top to prevent electrolyte from flowing out.
Check whether the sensor is damaged
(glass broken?).
Remove the watering cap from the sensor tip
and rinse the sensor with water.
1. Before installing the sensor, make sure
that the retractable fitting is in
SERVICE
position
(see page 17 for information on
installing or removing a sensor).
2. Loosen the small coupling nut
(R)
– do not
remove it.
3. Unscrew and remove the large coupling
nut
(S)
and pull the detached sensor holder
upwards.
4. Insert the sensor
(A)
through the gaskets
and Teflon washers.
Push the sensor downwards.
Carefully continue pushing the sensor
against the resistance of the gasket in the
immersion tube until it reaches the stop
position.
5. Replace the sensor holder you have
detached in step 3.
Hand-tighten the large coupling nut
(S)
and
then the small coupling nut
(R)
.
6. Connect the air pressure for the sensor
pressure chamber to the NW 6 connection
nipple
(AL)
.
7. Connect the cable jack.
Hold the cable in a loop and fix it using
clamp
(E)
.
Note:
The cable loop must be long enough
so that the cable does not impede the
stroke movement of the retractable fitting.
8. Connect the equipotential bonding cable to
terminal
(F)
(if required).
R
S
W
A
AL
Conditions:
• Sensor: 250 mm, Ø 12 mm, e.g., Knick SE 551
• Air pressure in the sensor pressure chamber: 0.5 to 1 bar above that of the process medium to
ensure that the electrolyte flows from the reference electrode to the process medium
Installing the Sensor