Installing the Sensor
Before installing the sensor, make sure that the
retractable fitting is in
SERVICE position
.
1. Use appropriate sensors
(A)
only:
Diameter: 12 mm Length: 225 mm
Observe the pressure resistance of the sensor.
2. Make sure that the slide washer
(C)
and the
O-ring
(D)
are correctly positioned and not
damaged.
3. Insert the sensor and screw in the sensor head
(J)
(19 mm A/F, PG 13.5 thread) with a max.
torque of 3 Nm (recommended tool: 19 mm
mounting wrench, e.g., Knick ZU 0647).
Note:
When tightening the sensor, you must
overcome the elastic force of the internal
sensor monitoring.
4. Connect the cable jack
(G)
.
Hold the cable in a loop and fix it using clamp
(E)
.
Note:
The cable loop must be long enough
so that the cable does not impede the stroke
movement of the retractable fitting.
5. Connect the equipotential bonding cable to
terminal
(F)
(if required).
6. Mount the protective cap (ZU 0759) if required
(see user manual of protective cap).