MemoTrans
41
No. Message
S
1)
D
2)
F
3)
Sensor type Configurable Tests or remedial action
241 Firmware error
F
On
On
All
No
Software error - internal
1. Carry out software update.
2. Replace backplane.
3. Please contact the Service Department
and quote the number shown.
243 Firmware error
F
On
On
All
No
Software error - internal
1. Carry out software update.
2. Replace backplane.
3. Please contact the Service Department
and quote the number shown.
284 Firmware update
F
On
On
All
No
Firmware update active, please wait.
285 Update error
F
On
On
All
No
Firmware update failed. Possible reasons:
Read error, SD card is faulty, incorrect
firmware on SD card.
1. Check device settings.
2. Check electronics module type.
373 Temperature of elec-
tronics is high
M
On
Off
All
No
Temperature of backplane electronics is
high.
• Check ambient temperature and
energy consumption.
384 Firmware error
F
On
On
All
No
Sensor measuring sequence counter.
No measurement signal from sensor.
1. Update software.
2. Contact the Service Department.
408 Calibration aborted
M
Off
Off
P, C, O,
No
Calibration aborted.
411 Upload/download
active, please wait
C
On
Off
All
No
Up-/Download active,
• please wait.
460 Output undershot
S
On
Off
All
No
Current output undershot.
Measured value out of specified current
range.
Possible reasons: Sensor / sample line in
air, air pockets in the fitting, incorrect
flow to sensor, sensor / sample line is
contaminated.
1. Check sensor.
2. Check application.
3. Clean sensor / sample line.
461 Output overshot
S
On
Off
All
No
Current output overshot. Measured value
out of specified current range.
Possible reasons: Sensor / sample line in
air, air pockets in the fitting, incorrect
flow to sensor, sensor / sample line is
contaminated.
1. Check sensor.
2. Check application.
3. Clean sensor / sample line.
1)
Status signal factory setting
2)
Diagnostics message factory setting
3)
Fault current factory setting