51
Nominal Operating Conditions
Ambient temperature
–20
…
85 °C / -4
…
185 °F
Transport / storage
temperature
-40
…
85 °C / -40
…
185 °F
Process temperature
Fitting in measuring position:
T(process) = max. 100 °C / 212 °F, continuous operation
T(ambient) = max. 60 °C / 140 °F, continuous operation
Fitting in service position
T(process) = max. 145 °C / 293 °F, continuous operation
T(ambient) = max. 60 °C / 140 °F, continuous operation
Relative humidity
5
…
95 % not condensing
Max. altitude above MSL
< 2000 m / < 6562 ft.
Output
4
…
20 mA current loop floating, protected against inverse polarity
Linearization/transmission
behavior
Linear
Supply voltage
12.6
…
30 V DC (with a fault current > 20 mA)
14
…
30 V DC (with a fault current < 4 mA)
Surge protection
IEC 61 000-4-4 and IEC 61 000-4-5, ± 1 kV each
Signal on alarm
3.6
…
23 mA
Connection
2-wire cable
4
…
20 mA positive: blue
4
…
20 mA negative: white
Housing
PEEK, color: light gray, RAL 7035
Memosens closure
PEEK, color: black
Cable
TPE, color: black, Ø approx. 5 mm
Optical waveguide
PC; color: transparent
Cable length
3 m / 10 ft; 7 m / 23 ft; 15 m / 49 ft
Impact loads
The product is designed for mechanical impact loads of 1 J (IK06) as per the
requirements of EN 61010-1.
Dimensions
See dimension drawing
Protection
IP67, IP68, NEMA 6
Weight
MemoTrans:
with 3 m / 10 ft cable approx. 190 g / 7 oz
with 7 m / 23 ft cable approx. 380 g / 13 oz
with 15 m / 49 ft cable approx. 760 g / 27 oz
Connections
Terminals, conductor cross-section max. 2.5 mm
2
Simulation
Specific parameters can be simulated for test purposes:
Current, measured value, or temperature