MemoTrans
27
11.3 Adapting Sensor Diagnostic Information
This menu branch is used for specifying warning limits, and for defining whether and how diagnostics
tools should be used.
11.3.1 Impedance Monitoring
Path:
Application
Sensor
Extended setup
Diagnostic settings
Function
Options
Info
Glass impedance
Upper limit
Selection
• Off
• On
Factory setting
On
On:
The Sensor Check System (SCS) operates with the following
settings for the upper warning and alarm limits.
Off:
Monitoring of the upper warning and alarm limits is switched off.
Upper alarm limit
0
…
10000 MΩ
Factory setting
3000 MΩ
Diagnostics code and associated message text:
124
Sensor glass
Upper warning limit
0
…
10000 MΩ
Factory setting
2500 MΩ
Diagnostics code and associated message text:
125
Sensor glass
Lower limit
Selection
• Off
• On
Factory setting
Off
On
: The Sensor Check System (SCS) operates with the following
settings for the lower warning and alarm limits.
Off:
Monitoring of the lower warning and alarm limits is switched off.
Lower warning limit
0
…
10000 kΩ
Factory setting
100 MΩ
Diagnostics code and associated message text:
123
Sensor glass
Lower alarm limit
0
…
10000 kΩ
Factory setting
0.0 MΩ
Diagnostics code and associated message text:
122
Sensor glass
Reference impedance
Selection
• Off
• On
Factory setting
On
Upper alarm limit
0
…
1000 kΩ
Factory setting
100.0 MΩ
Upper warning limit
0
…
1000 kΩ
Factory setting
50.0 MΩ