MemoTrans
43
No. Message
S
1)
D
2)
F
3)
Sensor type Configurable Tests or remedial action
513 Zero warning
M
On
Off
O (amp.)
No
Zero point warning. Measurement can
still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
515 Sensor calibration
M
On
Off
P (ISFET)
No
Max. operating point warning.
Measurement can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
517 Sensor calibration
M
On
Off
P (ISFET)
No
Min. operating point warning.
Measurement can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
518 Sensor calibration
M
On
Off
P, O
No
Delta slope warning. Measurement can
still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
520 Sensor calibration
M
On
Off
P, O
No
Delta zero point warning. Measurement
can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
522 Sensor calibration
M
On
Off
P (ISFET)
No
Delta operating point warning.
Measurement can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
532 License error
M
On
Off
All
No
License error
1)
Status signal factory setting
2)
Diagnostics message factory setting
3)
Fault current factory setting