11.3.12 Pharmaceutical Water
Path:
Application
Sensor
Extended setup
Diagnostic settings
Function
Options
Info
Pharmaceutical water
Here you can make settings for monitoring pharmaceutical water
in accordance with the United States Pharmacopoeia (USP) or
European Pharmacopoeia (EP).
The uncompensated conductivity value and the temperature are
measured for the limit functions. The measured values are
compared against the tables defined in the standards. An alarm is
triggered if the limit value is exceeded. Furthermore, you can also
set a preliminary alarm (warning limit) which signals undesired
operating states before they occur.
Function
Selection
• Off
• EP
• USP
Factory setting
Off
The alarm values are stored in the device in accordance with
USP <645> or EP <169> specifications.
You define the warning limit as a % of the alarm value.
Off
10.0
…
99.9 %
Factory setting
80.0 %
Diagnostics code and associated message text:
915
USP / EP warning
If the value exceeds the USP or EP alarm values saved in the
software, diagnostics message 914
USP/ EP alarm
is displayed.
11.3.13 Cap Calibrations
Path:
Application
Sensor
Extended setup
Diagnostic settings
Function
Options
Info
No. calibrations cap
Not displayed for optical oxygen sensors.
The calibration counters in the sensor make a distinction between
sensor calibrations and calibrations with the membrane cap cur-
rently used. If this cap is replaced, only the (cap) counter is reset.
Function
Selection
• Off
• On
Factory setting
Off
Specify how many calibrations may be performed with a membrane
cap before the cap has to be replaced. The number depends heavily
on the process and must be determined individually.
Warning limit
1
…
1000
Factory setting
6
Diagnostics code and associated message text:
535
Sensor check