11.3.3 Slope
pH, pH/ORP, oxygen
Path:
Application
Sensor
Extended setup
Diagnostic settings
Function
Options
Info
pH
Slope
The slope characterizes the sensor condition. The greater the deviation
from the ideal value (pH), the worse the condition of the sensor.
Warning limit
25.00
…
65.00 mV/
pH
Factory setting
55.00 mV/pH
Specify your limit values for slope monitoring. Diagnostics code and
associated message text:
509
Sensor calibration
Oxygen
Upper warning limit
0
…
200.0 %
Factory setting
140.0 %
Diagnostics code and associated message text:
511
Sensor calibration
Lower warning limit
0
…
200.0 %
Factory setting
60.0 %
Diagnostics code and associated message text:
509
Sensor calibration
11.3.4 Delta Slope
pH, pH/ORP, oxygen
Path:
Application
Sensor
Extended setup
Diagnostic settings
Function
Options
Info
pH and pH/ORP combo sensors
Delta slope
The device determines the difference in slope between the last calibra-
tion and the penultimate calibration, and issues a warning or an alarm
depending on the setting configured. The difference is an indicator for
the condition of the sensor. The greater the change, the greater the
wear experienced by the pH-sensitive glass membrane as a result of
chemical corrosion or abrasion.
Function
Selection
• Off
• On
Factory setting
Off
Switches the function on or off
Warning limit
0.10
…
10.00 mV/pH
Factory setting
6.00 mV/pH
Specify your limit values for monitoring the slope differential.
Diagnostics code and associated message text:
518
Sensor calibration
Oxygen
Delta slope
The device determines the difference in slope between the last calibra-
tion and the penultimate calibration, and issues a warning or an alarm
depending on the setting configured. The difference is an indicator for
the condition of the sensor.
An increasing change indicates the formation of buildup on the sensor
diaphragm or electrolyte contamination. Replace the diaphragm and
electrolyte as specified in the instructions in the sensor User Manual.