MemoTrans
40
No. Message
S
1)
D
2)
F
3)
Sensor type Configurable Tests or remedial action
189 Operating time
M
On
Off
O
No
Operating hours > 5 °C, measurement can
still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
191 Operating time
M
On
Off
O
No
Operating hours > 30 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
192 Operating time
M
On
Off
O
No
Operating hours > 40 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
193 Operating time
M
On
Off
P, C, O
No
Operating hours > 80 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
194 Operating time
M
On
Off
P
No
Operating hours > 100 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
195 Operating time
M
On
Off
C
No
Operating hours > 120 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
197 Operating time
M
On
Off
C
No
Operating hours > 140 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
198 Operating time
M
On
Off
C
No
Operating hours > 150 °C, measurement
can still take place.
1. Replace sensor.
2. Change monitoring limit.
3. Disable monitoring.
199 Operating time
M
On
Off
All
No
Total operating hours
202 Self-test active
F
On
On
All
No
Self-test active, please wait.
215 Simulation active
C
On
Off
All
No
Simulation active. End simulation by
changing to measuring mode.
216 Hold active
C
On
Off
All
No
HOLD active. Output values and channel
status are on hold.
• Please wait.
1)
Status signal factory setting
2)
Diagnostics message factory setting
3)
Fault current factory setting