MemoTrans
21
Function
Options
Info
SV value
TV value
QV value
pH, ISFET, ORP, and pH/ORP
units
• pH
• Raw value pH, mV(pH)
• Glass impedance
• Impedance reference
• rH
• ISFET leakage current
• ORP mV
• ORP %
• Temperature
Oxygen units
• Partial pressure
• % saturation
• Concentration (liquid) unit
• Conc. (gaseous) unit
• Raw value nA
• Temperature
Conductivity units
• Conductivity
• Resistance
• Raw value (cond. uncomp.)
• Temperature
Protocol-specific data, dynamic variables of HART
communication.
SV = secondary
TV = tertiary
QV = quaternary
8.4.5 Hold
The hold state is a safe condition during configuration and calibration.
Path:
System
Hold
Function
Options
Info
Hold release time
0
…
600 s
Factory setting
0 s
The hold status is maintained for the duration of the
delay time when you switch to the measuring mode.
Hold behavior
Selection
• Ignore
• Freeze
• Fixed value
Factory setting
Freeze
Hold current
3.6
…
23 mA
Factory setting
0 s
Hold current setting only possible if hold behavior
selection is “Fixed value”.
Manual hold
Selection
• Off
• On
Factory setting
Off
On: You can use this function to set the channel
manually to “Hold”.
Off: No channel-specific hold