No. Message
S
1)
D
2)
F
3)
Sensor type Configurable Tests or remedial action
500 Sensor calibration
M
On
Off
All
No
Calibration aborted, main measured
value varies.
Reasons: sensor too old, sensor occasion-
ally dry, calibration value not constant.
1. Check sensor.
2. Check calibration solution.
501 Sensor calibration
M
On
Off
All
No
Calibration aborted, temperature
measured value varies.
Reasons: sensor too old, sensor
occasionally dry, temperature of
calibration solution not constant.
1. Check sensor.
2. Regulate calibration solution
temperature.
505 Sensor calibration
M
On
Off
P, O
No
Max. zero point warning. Measurement
can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated.
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
507 Sensor calibration
M
On
Off
P, O
No
Min. zero point warning. Measurement
can still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated.
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
509 Sensor calibration
M
On
Off
P, O
No
Min. slope warning. Measurement can
still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated.
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.
511 Sensor calibration
M
On
Off
P, O
No
Max. slope warning. Measurement can
still take place.
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated.
1. Check or replace sensor.
2. Check or replace calibration solution.
3. Repeat calibration.