MemoTrans
16
8.4.2 Configuring the Sensor
Path:
Application
Function
Options
Info
Sensor
Sensor-dependent settings
Sensor type
Display the sensor type
Order code
Order code of sensor
ORP
Display % value
Display ORP as percentage
(pH/ORP combo sensors only)
Damping
The damping causes a floating
average curve of the measured
values over the time specified.
Damping pH,
Damping ORP,
Damping oxygen,
Damping conductivity
0
…
60 s
Factory setting
0 s
These functions determine the
damping of the primary value of
the connected sensor.
Damping temperature
0
…
60 s
Factory setting
0 s
These functions determine the
damping of the integrated
temperature sensor.
TAG control
Sensor check
Selection
• Off
• TAG
• Group
Factory setting
Off
Group
Range: 0
…
65535
Extended setup
Conductivity:
Current cell
constant
Read only
Value currently saved in the sensor.
Compensation
Selection
• None
• Linear
• NaCl (IEC 746-3)
• Water ISO7888 (20 °C)
• Water ISO7888 (25 °C)
• Ultrapure water (NaCl)
• Ultrapure water (HCl)
Factory setting
Linear
Various methods are available to
compensate for the temperature
dependency.
Depending on your process, decide
which type of compensation you
want to use.
Alternatively, you can also select
None
and thus measure
uncompensated conductivity.
Compensation for ultrapure water is
only available with contacting
sensors.
Meas. ref. temp.
-5.0
…
100.0 °C
(23.0
…
212.0 °F)
Factory setting
25.0 °C (77.0 °F)
Reference temperature for calculating
the temperature-compensated
conductivity
Factor alpha
0.000
…
20.000 %/K
Factory setting
2.100 %/K
Enter the conductivity coefficient of
your process medium