![KLS Martin IDP 101 S air plus Скачать руководство пользователя страница 91](http://html.mh-extra.com/html/kls-martin/idp-101-s-air-plus/idp-101-s-air-plus_mounting-instructions_1978996091.webp)
Suspension Arm Systems Independant
Revision 1
91
8.4
Adjusting the Vertical Lift on the Spring Arm
When replacing an end device (e.g. TFT monitor or medical device), the spring arm must be
adjusted in the horizontal position (0°).
Opening the upper casing on the front
•
Observe the safety notice, see section 8.1
“Safety Notice for Adjustment Work”,
page 86.
•
Successively press a suitable screwdriver
into the 2 openings and unlock the
2 latches .
•
Fold the upper cover on the front
upwards until it latches.
Adjusting the Vertical Lift
Tools to be used:
Hexagon spanner SW 10 and ring spanner
SW 18
•
Loosen hexagonal nut M12
ISO 4035 and
turn it back.
•
Insert hexagon spanner into adjustment
screw .
To reduce the vertical lift:
•
Turn hexagon spanner to the left
(counterclockwise) as shown in the
illustration.
To increase the vertical lift:
•
Turn hexagon spanner to the right
(clockwise) as shown in the illustration.
•
Perform a function test.
•
Tighten hexagonal nut .
Fig. 8-4:
Adjusting the vertical lift on the spring arm