Suspension Arm Systems Independant
74
Revision 1
•
Feed all power cables through the tension relief and connect them to the terminal
unit according to the wiring diagram provided by the customer.
Ensure that the power cables are safely in place in the tension relief :
•
All power cables must sit securely in the tension relief .
7.7.5
Checking the Supply Cables
•
Carefully check that the supply cables are not pinched or bent through the entire range of
rotation of the extension arm and the spring arm.
7.8
Checks
7.8.1
Gas Type Check (M6 Service Head only)
NOTICE
The gas type check must be performed and documented by the supplier of the central gas
system or a qualified specialized company.
The following checks must be performed:
•
Connections and labeling acc. to DIN EN ISO 91701 or DIN EN ISO 91702
•
Leakage acc. to DIN EN ISO 11197
•
Obstruction acc. to DIN EN ISO 73961 or DIN EN ISO 73962
•
Solid contaminants acc. to DIN EN ISO 73961 or DIN EN ISO 73962
•
Gas type acc. to DIN EN ISO 73961 or DIN EN ISO 73962
7.8.2
Mechanical Collision Check
Ensure that there are no collisions when swiveling the suspension arm system. For example
with:
•
other suspension arm systems
•
ceilings or walls
•
other components