![KLS Martin IDP 101 S air plus Скачать руководство пользователя страница 67](http://html.mh-extra.com/html/kls-martin/idp-101-s-air-plus/idp-101-s-air-plus_mounting-instructions_1978996067.webp)
Suspension Arm Systems Independant
Revision 1
67
Installing and dismantling brake pipes
The pneumatic supply pipe must be pressureless:
•
To install the brake pipes, push them into the brake connecting point. If the brake pipe is
correctly positioned, it can no longer be withdrawn from the brake connecting point.
•
To disengage the brake pipes on the plug connector push in the unlocking
mechanism and then remove the brake pipes.
Laying and Connecting the pneumatic brake pipes
•
Insert the pneumatic brake lines into the mounting opening in the in the bearing
support of the spring arm and lay them above the swivel axis (swivel joint ).
•
Plug the air supply hose (black marking) onto the plug connector (black marking).
•
Plug the brake pipe (green marking) onto the plug connector (green marking).
•
Plug the brake pipe (blue marking) onto the plug connector (blue marking).
•
Connect the air supply hose and the 2 brake pipes to the plug connector on the
shelf as described in the Mounting Instructions of the M6 Service Head.
The brake pipes can be too long depending on the installation position of the shelf:
•
Cut the brake pipes to length.
Carry out a functional test after mounting the M6 Service Head:
•
Push the green brake button
.
The brake on the upper extension arm must release.