![KLS Martin IDP 101 S air plus Скачать руководство пользователя страница 69](http://html.mh-extra.com/html/kls-martin/idp-101-s-air-plus/idp-101-s-air-plus_mounting-instructions_1978996069.webp)
Suspension Arm Systems Independant
Revision 1
69
7.6.10
Feeding the Supply Cables Through the Spring Arm, Single Arm, to the
M6 Service Head
•
Observe the safety notes, see section 7.6.1 “Safety Notes”, page 58.
Fig. 7-29:
Feeding the supply cables through the spring arm, single arm, to the M6 Service Head
The power cables, pneumatic pipes, earthing and control cables, as well as the gas hoses are
pre-installed in the M6 Service Head and must be laid through the pendant system. Dedicated
cables, e.g. for nurse call or telephony, must be laid separately through the suspension arm
system.
•
Carefully lay the supply cables and supply tubes through the suspension arm system to
the interface plate.
•
Lay the supply cables and supply tubes
through the spring arm above the swivel axis
(swivel joint ) and lead them into the mounting opening in the bearing block of the
spring arm:
—
Then readjust the M6 service head without pulling on the supply cables and supply
tubes . Ensure that cables that have already been laid are not damaged or torn off.
•
Lead dedicated cables, e.g. for nurse call or telephony, through the suspension arm
system.
•
Mount the M6 service head to the suspension arm system according to the Mounting
Instructions of the M6 service head.