84
Keysight NFA Series Noise Figure Analyzers Service Guide
Instrument Messages
Event Messages
–314
Save/recall memory
lost
The nonvolatile data saved by the *SAV? command has been lost
–315
Configuration memory
lost
The nonvolatile configuration data saved by the instrument has been lost
–320
Storage fault;
A problem was found while using data storage. The error is not an indication of
physical damage or failure of any mass storage element
–321
Out of memory
An internal operation needed more memory than was available. Report this
error to the nearest Keysight Technologies sales or service office
–330
Self-test failed
A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office
–340
Calibration failed
The instrument requires an Align
All
Now. Restore the alignment by pressing
System, Alignments, Align
All
Now
–340
Calibration failed;
<failure msg>
<port>
The calibration for one of the I-Q ports did not succeed. The information in the
“failure msg” field can be used to troubleshoot this problem. Contact Keysight
technical support
–350
Queue overflow
An error occurred that did not get put in the error queue because the queue
was full
–360
Communication error
There was a problem with instrument remote communications. The exact
problem cannot be specifically identified
–360
Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
The Keysight Smart Noise Source connected to the instrument has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or another
hardware fault exists. Check SNS on another instrument, NFA and ESA are also
SNS compatible instruments. Check that the device is not an Keysight power
sensor which uses the same cable interface
–360
Communication error;
SNS is not connected
The Keysight Smart Noise Source connected to the instrument has failed to be
read by the application. Please disconnect and reconnect the SNS. If this
continues to fail, then the SNS may have had its EEPROM corrupted or another
hardware fault exists. Check SNS on another instrument, NFA and ESA are also
SNS compatible instruments. Check that the device is not an Keysight power
sensor which uses the same cable interface
–361
Parity error in
program message
A parity bit was not correct when the data was received. For example, on a
parallel port
–362
Framing error in
program message
A stop bit was not detected when data was received. For example, on a remote
bus port
–363
Input buffer overrun
A software or hardware input buffer on a port overflowed with data because of
improper or nonexistent pacing
–365
Time out error
There was a time-out problem in the instrument. The exact problem cannot be
specifically identified
Err#
Message
Verbose/Correction Information
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...