162
Keysight NFA Series Noise Figure Analyzers Service Guide
RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
Input Attenuator A Power Level Verification
Press
Mode Preset
,
Input/Output
,
RF Calibrator
,
50 MHz
,
AMPTD
,
Attenuation
,
Mech Atten, 0 dB
on the analyzer. Refer to
, remove
cable W11 from A9
(1)
Output. Measure the 50 MHz calibrator signal on the
output of the attenuator using a functioning Spectrum Analyzer.
Figure 4-13
Cable W11