Keysight NFA Series Noise Figure Analyzers Service Guide
65
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
E Cal Path System Gain Algorithm
Measures the amplitude error of the low band signal
path by injecting the 50 MHz E-Cal signal into the A13
Front End assembly. The E-Cal signal originates on the
A16 Reference assembly.
Signal path from A13
Front End to A3 Digital
IF
A16 Reference may not be
providing 50 MHz E-Cal
Low Band Nominal Path System Gain
Algorithm
Aligns low band (3 Hz to 3.6 GHz path) with preamp
OFF. Dependent on successful E-Cal alignment.
Input attenuators set to 10 dB and DC coupled.
A13 Front End
50 MHz calibrator on A16
Reference.
A9 and A10 Input attenuators
Low Band Preamp Path System Gain Algorithm
Aligns low band (3 Hz to 3.6 GHz path) with preamp ON.
Input attenuator set to 20 dB.
(Options Pxx)
A13 Front End
50 MHz calibrator on A16
Reference.
A9 and A10 Input attenuators
Preselector Two Point Tuning Algorithm
Aligns the YTF preselector centering near the maximum
and minimum frequency points of the YTF.
Uses the high band preamp noise as stimulus signal.
A12 YTF Preselector
or
A15 Front End
Controller
or
All Low Band
A13 Front End
High Band Nominal Path System Gain
Algorithm
Anything in Band 1
signal path
4800 MHz calibrator on A16
Reference
High Band Preamp Path System Gain
Algorithm
Measures change in signal level with high band preamp
turned ON and OFF.
Input attenuator set to 30 dB.
(Options P07, P13, P26, P32, P44)
A11 Low Band Switch
4800 MHz calibrator on A16
Reference
A9 and A10 Input attenuators
A15 Front End Controller
Burst Carrier Trigger Curve Fit Algorithm
Detects the level of the signal envelope before filtering
at the 22.5 MHz IF, and compares that detected level
against the trigger threshold DC level.
A2 Analog IF
Trigger Interpolation
A3 Digital IF
DIF 25 Pulse Stretcher Alignment Algorithm
A3 Digital IF
Table 2-1
Initial Alignments
Alignment Description
Most Probable
Hard ware Failure
Related Hard ware
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...