Keysight NFA Series Noise Figure Analyzers Service Guide
85
Instrument Messages
Event Messages
–221 Settings Conflict Errors
This is one of the errors in the standard SCPI error range –200 to –299. For all
other errors in this range, see
“–200 to –299, Execution Errors” on page 96
.
Err#
Message
Verbose/Correction Information
–221
De-emphasis only
available in FM
The de-emphasis function is only available if FM demod is selected
–221
Function not
available in Zero
Span
The function you are trying to access is not available in zero span
–221
Setting conflict;
<trigger source>
trigger is not
available while input
is <input port>
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not available with the
current input port (RF, IQ, etc.)
–221
Setting conflict;
Differential setting
determined by probe
type
A probe is connected that has a built in Differential setting. The setting
cannot be changed manually
–221
Setting conflict;
Input Z unavailable
when probe sensed
A probe is connected and the Input Z is set based on the probe type. It
cannot be changed manually
–221
Settings conflict;
A legal command was received but it could not be executed due to the
current device state
–221
Settings conflict;
*.CSV file format is
not available in this
measurement.
You cannot load or save base instrument traces, as this is not supported by
the Log Plot measurement
–221
Settings conflict; <Q
Param> cannot be
changed when Q same
as I
When the
Q Same as I
parameter is set to Yes, the I parameter value is
copied to <Q Param> and the <Q Param> value cannot be changed. Set
Q
Same as I
to No to enable explicit control of the <Q Param> value
–221
Settings conflict; A
Valid User Cal is
required. Optimize
aborted
Optimize Preselector can only be performed if a valid user cal exists and is
applied to current results. Perform a user cal first or apply existing cal
–221
Settings conflict;
Auto Tune not
available in Tracking
Source mode
The Auto Tune feature cannot be used when you are using a Tracking Source
–221
Settings conflict;
BTS gain is not
available in this
Mode
Base Transceiver Station gain correction is not available in some Modes, or in
some measurements (for example, the SA measurement)
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...