Keysight NFA Series Noise Figure Analyzers Service Guide
Assembly Replacement Procedures
A5 Solid State Drive
2.
. Remove the four screws
(5)
(0515-1035) from the
disk drive tray that secure the SSD
(1)
.
Figure 15-36
Disk Drive Tray Screws
Replacement
1.
. Place the new SSD into the tray assembly and
attach with the four screws
(5)
(0515-1035). Torque to 9 inch-pounds.
2.
. Slide the SSD into the CPU assembly and push to
mate the connector. Secure the thumb screw to 9 inch-pounds.
Make sure not to use the 0515-0372 screw because for the PC6 processor
it will cause the SSD to interfere with the CPU memory card.
Chapter 16, “Post-Repair Procedures.”
and perform all of the
specified tasks for replacing the A5 Disk Drive assembly.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...