Keysight NFA Series Noise Figure Analyzers Service Guide
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
No Keysight Splash Screen Displayed
(Black background with white “Keysight Technologies” text)
A problem of not displaying the Keysight splash screen could be caused by
many different things. It could be due to a down power supply, a processor
hardware problem, an instrument boot-up process error, a display section
failure, etc.
This procedure assumes that the green Power On LED on the front panel does
turn on when the instrument is turned on. If it doesn't, refer to the
Power On LED Does Not Illuminate”
section before proceeding.
1.
Remove the AC power cord and then remove the instrument cover. Refer
Chapter 15, “Assembly Replacement Procedures”, on page 389
in this
manual.
2.
Remove the MP7 Top Brace. Refer to
Chapter 15, “Assembly Replacement
in this manual.
3.
With the AC power applied and the On/Off button turned on, verify that all
of the power supply voltages are at their proper level. This can easily be
done by viewing the power supply LEDs on the back side of the A7
Midplane board. See
“A7 Midplane Assembly Troubleshooting” on
of this manual.
Are all of the power supply voltages at the proper level?
If yes:
Proceed to
.
If no:
After verifying that the connections between the A6 Power Supply,
the A7 Midplane board, and the A8 Motherboard are all
mechanically and electrically secure, replace the A6 Power Supply
assembly.
4.
Connect an external VGA monitor to the rear panel display output and
cycle power on the analyzer.
Does the external monitor display the correct information?
If yes:
Proceed to the
“Troubleshooting a Blank Display”
section in this
chapter.
If no:
Replace the A4 CPU Board assembly.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...