Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
If there is a problem with any of these initializations not completing or causing
an error message to be displayed refer to the instrument Event Log. This can
be accessed by using an external USB keyboard and mouse and selecting
Start, Command Prompt, enter Eventvwr.exe, and select OK.
Once the Event Viewer comes up, look under SA for the latest error entries.
Double-click on the entries to view further details, which should give you some
idea of what the problem is.
Fails an Initial Alignment
Troubleshooting Alignment Failures Using the Alignment History Text
File
At instrument power on, an initial alignment is automatically performed. It is
also possible to manually trigger an alignment by pressing
System
,
Alignments
,
Align Now
,
Align Now All
. The power on alignment performs a
few more alignments than the manual Align Now All routine.
The number of alignments depends on the hardware options present in the
signal analyzer.
Using an external USB keyboard and mouse, access the alignment history,
which can be found in a text file at:
E:AlignDataStorage\AlignmentHistory.txt
You will find the most current alignment information at the bottom of this file.
Look for the failed alignments. From the windows task bar you can click Edit,
Find and type in FAILED and search for a failure.
Note that the information in the history file usually shows that multiple
measurements are made for a given alignment routine, and measured values
with upper and lower limits are given. For example, the Mechanical Attenuator
Algorithm measures and displays all of the attenuator step values. In this case,
you can view which steps fail, and by how much. Then based on the input
attenuator drawing on the RF Block Diagram, you can determine which
attenuator steps are available on each of the two input attenuators. Then you
can view the troubleshooting information for the Front End Controller or
deduce from the attenuator steps that fail, which attenuator is faulty.
Assure there is no 50 MHz signal above 0 dBm applied to the signal
analyzer input when the alignment routine is performed. This can cause
errors.
The description of the alignment and the alignment hardware
dependencies are listed in
Table 2-1, “Initial Alignments,”
.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...