RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
Press
Mech Atten
and enter 10 dB. The 50 MHz calibrator signal measured on
the functioning Spectrum Analyzer should measure 10 dB lower than the first
step (~
−
35 dBm). Continue testing all the attenuation settings as described in
.
Table 4-10
Attenuator Setting
(dB)
Expected Level
(dBm @ 50 MHz)
Tolerance
16
−
41
±
3 dB
20
−
45
±
3 dB
30
−
55
±
3 dB
40
−
65
±
4 dB
50
−
75
±
4 dB
60
−
85
±
4 dB