121
Instrument Messages
Condition Messages
217
7
Demod Error
E
This error is normally generated because of one of the
following reasons:
1.
There is no carrier signal.
2.
Walsh channels other than the pilot are
active.
3.
There is some other modulation problem
that will prevent the measurement from
being made.
This problem must be corrected before the
measurement can continue.
cdma 2000 & W-CDMA: Cannot correlate to the input
signal and no active channel is found. (from composite
EVM measurement) An active channel must meet the
default threshold criteria that it is within 20
dB of the
highest power code channel. The threshold can be
changed using the active set threshold function in the
Meas Setup menu.
cdmaOne: A correlation failure with the pilot CDMA
channel occurred during synchronous demodulation.
1xEV-DO: Cannot correlate to the input signal and no
active channel is found. (from composite EVM
measurement) An active channel must meet the default
threshold criteria that it is within 20
dB of the highest
power code channel. The threshold can be changed
using the active set threshold function in the Meas
Setup menu.
In PSA, this error was reported as one of the following
error numbers: 10872, 10962, 13070, 10228, 10768
217
7
Demod Error;Can’t
correlate
E
Cannot correlate to the input signal and no active
channel is found. (from composite EVM measurement)
An active channel must meet the default threshold
criteria that it is within 20
dB of the highest power code
channel. The threshold can be changed using the active
set threshold function in the Meas Setup menu.
217
7
Demod Error;Data
interval too
short
E
There are not enough input I/Q pairs for the
measurement calculation. This may be caused by an
incorrect data capture.
217
7
Demod Error;No
active channel
E
There is no active channel detected.
217
7
Demod Error;Not
an active slot
E
There is no active slot detected.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...